Emerson Rosemount 370XA System Reference Manual page 143

Gas chromatograph
Hide thumbs Also See for Rosemount 370XA:
Table of Contents

Advertisement

System Reference Manual
7P00370-H01
9.
10.
6 - 6
Reconnect the connector to the solenoid, ensuring the
retention clip engages. The connectors are labeled according
to the following table:
Valve label
Manifold label
AV1
AV1
AV2
AV2
AV3
AV3
Reinstall the module on the 370XA and run an analysis to
confirm correct operation. See
must test the module and adjust the analytical method (valve
timings, integration events, and calibration settings). See "Auto
valve timing" on page 4-37." on page
370XA Gas Chromatograph
Description
Sample valve
Back-flush valve
Dual column valve
"After you make the repairs, you
6-50.
APRIL 2014