APRIL 2014
Maintenance Mode
Switches the GC to maintenance mode and triggers an alarm that the
GC is down for maintenance. This option can be switched on or off by
the user.
Enable CV Check
Std Comp Table Version
Indicates which version of the GPA's standard component table is
being used.
Default Stream Sequence
Displays the default sequence to be used by the GC during auto-
sequencing.
4.3.3.2

The Component Data Table screen

Component
Displays the list of available components for the selected stream.
Ret Time
A component's retention time, which is the time, in seconds, when the
apex of the component's peak is expected to appear.
Resp Fact
A component's response factor is equal to the value of the
component's peak divided by the component's concentration value.
Calib Conc
The concentration amount, in mole percent, of the component that is
present in the calibration gas.
System Reference Manual
7P00370-H01
4 - 21