Parts required
Tools required
To maximize the time between repairs to the module, it is
recommended to overhaul all three analytical valves and the sample
shut-off valve at one time, rather than repairing only a single valve
when it is suspect.
Perform this procedure in a clean environment. When removing the
components from the valve, place them onto a clean work surface to
avoid contaminating the valve components.
This procedure assumes that you have removed the analytical
module from the GC.
1.
6 - 8
A valve overhaul kit for each valve to be serviced.
• A vise with aluminum soft-jaws or some other protection
to avoid marking the analytical valves with the vice jaws.
• A 2.5mm hex or allen wrench.
• An 11mm (7/16in) socket and a socket wrench.
• A torque wrench.
• A no-residue, evaporating electrical contact cleaner.
Remove the heater cap from the module.
370XA Gas Chromatograph
APRIL 2014