7P00370-H01
1.
2.
NOTICE
To ensure the proper and reliable operation of the GC, always use new o-rings
supplied as a part. O-rings should never be re-used or sourced from third-party
suppliers.
3.
4.
5.
6.
7.
5 - 6
Inspect the spring-loaded feed-through (1) to ensure all the o-
ringo-rings have been removed.
Install the new o-rings into the spring-loaded feed-through.
Inspect the under side of the new analytical module to ensure
that there are no o-rings stuck to the module base.
Connect the solenoid electrical connector to the male
connector leading to the solenoids.
Connect the male 18-pin signal connector to the connector on
the IMB circuit board.
Connect the two-pin heater connectors.
Align the groove on the module base to the front of the GC and
lower the module down onto the guide pins on the spring-
loaded feed-through.
370XA Gas Chromatograph
APRIL 2014