Compared with traditional sample preparations, the amount of mainte- nance of the PurCon is negligibly small. Due to the low dwell time of the sample in the PurCon, the preparation is carried out without any effect on existing component substances.
The feed is connected by a hose to the feed nozzle (PurCon housing base). The feed can be transported using an external pump or by using gravity to run it into the PurCon from a vessel placed higher than the PurCon.
Page 7
Permeate The permeate pump transports the permeate to the analyzers via a hose that is also supplied with the PurCon. It is connected to the per- meate nozzle. The quantity to be transported can be varied by a 4-step selection switch and, thus, optimally adapted to requirements.
Page 8
Introduction PurCon PurCon Permeat Power Fig. 4 Valve setting for emptying the sample preparation instrument PurCon Permeat Power Fig. 5 Valve setting for cleaning the sample preparation instrument...
In this case, wait for the instrument to adjust to room temperature before recommissioning. Caution Opening the instrument or adjustment, maintenance and repair work must only be performed by personnel authorized by WTW. EN 61010-1 The instrument complies with the requirements of EN 61010-1, protec- tion class I.
Page 10
Safety instructions PurCon of operation and secured against inadvertent operation. Safe operation is no longer possible if the instrument: l has been damaged in transport l has been stored under adverse conditions for a lengthy period of time l is visibly damaged l no longer operates as described.
PurCon Safety instructions Warning instructions and symbols Danger symbol according to the hazardous substance regulation that means: Causes severe corrosion. Do not allow this substance to get into the hands of children. Wear suit- able protective gloves and goggles/face mask when working with this substance.
PurCon sample preparation instrument Scope of delivery l Operating manual Caution The PurCon must be assembled by trained personnel only. The key that is supplied must be kept safe and must only be used by qualified electricians. D=8,4 mm...
PurCon (see chapter, T ECHNICAL Caution Ensure that the PurCon is installed in the immediate vicinity of the ana- lyzers to avoid dead times as the result of lengthy permeate feed routes. Caution The mounting stand must be fixed to the floor.
PurCon Commissioning Connecting the feed intake Connect the feed transport to the feed nozzle of the PurCon using a 3/4“ hose (ID 19 mm). Caution When setting up the quantity to be transported by the feed transport, follow the chapter T of this operating manual.
PurCon Installing the PurCon air outlet (option) We recommend to install the PurCon air outlet if the drain pipe is not entirely pressure-free. By using the air outlet, no overpressure can develop in the system which prevents the retentate from running off.
PurCon Commissioning Setting the manual valves The valves in the housing of the PurCon enable various operating modes. Caution Ensure that the setting is correct when commissioning the instrument (see Fig. 9). PurCon Permeat Power Fig. 9 Feed inlet open, outlet shut...
Use the power cable to connect the PurCon to an easily accessible mains socket. Warning The data on the nameplate (specified line voltage) of the PurCon must match that of the mains socket. See the chapter, T , for the connecting data.
PurCon Commissioning Inserting the pump hoses Proceed as follows: Take the pump cassette out of the holder by pressing the clamp strap. Fix the pump hoses supplied between the first and second stopper in the pump cassette. Replace the pump cassette in the holder and snap in the clamp strap.
Commissioning PurCon 3.10 Starting sample preparation The PurCon is now ready for operation. Before switching on the mains power, check the following points again: Are all the hose clamps used firmly tightened? Are the seals of the retentate nozzle seated correctly and tight?
PurCon Operation Operation The PurCon operates largely without the need for any maintenance. In normal operation, the instrument requires only a restricted number of maintenance activities. See the chapter M for more infor- AINTENANCE mation.
Displacing the pump tubings l Cleaning the air filter on the compressor l Replacing the pump tubings l Cleaning the PurCon The following table provides an overview of the maintenance activities. The intervals can vary, depending on the quality of the sample.
Maintenance PurCon Displacing the pump tubings The displacement of the pump tubings is necessary to prevent the mechanical wear of the pump tubings becoming too great. If this activity is carried out regularly, the lifetime of the pump tubings may exceed 6 months. The displacement can be carried out up to 5 times before the pump tubings become unusable.
PurCon Maintenance Replacing the pump tubings To replace the pump tubings, carry out the following steps: Switch off the permeate pump (bring the Permeate selector switch into position 0). Loosen the pump cassettes on the gripping lever. Take the pump cassettes out of the holder by pressing the clamp strap.
If the analyzers do not obtain sufficient permeate, they generate a cor- responding error message. In this case, the inside of the PurCon has to be cleaned. It is cleaned chemically by the effect of 1 % sodium hypochlorite solution. The cleaning solution can be produced from var- ious detergent concentrates by dilution.
Page 27
PurCon Maintenance On the following pages, cleaning the PurCon with hygienic cleaner and with detergent concentrate are described in detail. Method 1: To clean the PurCon, carry out the following steps: Cleaning using hygienic cleaner Disable the feed transport. (e. g. Dan Klorix) Switch off the permeate pump (selector switch to position 0).
Page 28
Maintenance PurCon Open the lid of the filler nozzle of the cleaning solution or unscrew the PurCon air outlet (option) on the top of the instru- ment. Cleaning the PurCon air outlet: – Unscrew the four lateral knurled-head screws and remove both caps and the inner insert.
Page 29
Subsequently, pour in 1.5 l hygienic cleaner and top up with water completely. Remove the funnel and screw the cap or the PurCon air outlet to the cleaning solution filler nozzle again. Allow the cleaning solution to act for approx. 30 min.
Page 30
Maintenance PurCon Method 2: To clean the PurCon, carry out the following steps: Cleaning using detergent concentrate Preparing the detergent (1% sodium hypochlorite): (13 % sodium Pour 300 ml of the detergent concentrate into a suitable vessel hypochlorite solution) (volume > 5l).
Page 31
After approx. 30 s, close the Outlet valve again. Open the lid of the filler nozzle of the cleaning solution or unscrew the PurCon air outlet (option) on the top of the instru- ment. Cleaning the PurCon air outlet: –...
In the case of drinking water or sensitive surface water, remove the out- let tube from the retentate nozzle. Remove the funnel and screw the cap or the PurCon air outlet to the cleaning solution filler nozzle again. Allow the cleaning solution to act for approx. 30 min.
Page 33
PurCon Maintenance To clean the filter, carry out the following steps: Switch off the PurCon at the power switch and pull out the mains plug. Unscrew the eight knurled-head screws on the top and remove the cover of the compressor.
Switch off the permeate pump on the Permeate selector switch. Switch the PurCon off on the power switch. Disconnect the PurCon from the mains by pulling out the mains plug. Using a suitable screw driver, open the fuse holder on the right side of the instrument.
What to do if ... Permeate is not being Is the permeate pump turning? If necessary, check the power supply of the PurCon. If the fault is not remedied, continue with step 2. Are the pump tubings properly mounted and functioning? Replace pump tubings if necessary.
Switch off the permeate pump (bring the permeate selector switch into position 0). Switch off the mains switch of PurCon. Fig. 20 Opening the filter unit Remove the eight hexsocket head screws on the top of the housing (pos.
Page 39
PurCon What to do if ... Fig. 21 Open filter unit Cut through the cable ties on the tube ends (pos. 3 in Fig. 21) and pull off the tubes. Make sure no dirt gets into the fittings. Pull out the filter plates using the clamp (pos. 4 in Fig. 21).
– 1 reducer for compressor connection – Cable ties Caution Maintenance work on the open PurCon may be carried out by trained specialists only. The key that is supplied must be kept safe and must only be used by qualified electricians. Before opening the operating panel of the instrument, switch it off at the mains switch and pull out the mains plug.
Page 41
Make sure the original tubes and the spare tubes have the same length. The correct lengths are given in Fig. 22. l Using cable ties, secure all Norprene tube connections at the filter unit and the two Y-pieces. Fig. 22 Permeate carrying tubes of the PurCon...
Page 42
Run the air tube in a loop so that its highest point is above the drain pipes of the filter unit (see Fig. 23). Otherwise, liquid may flow back into the air compressor when the PurCon is switched off. Fig. 23...
PurCon Technical data Technical data Environmental Storing temperature – 25 ... 60 °C conditions Operating temperature 0 ... 40 °C Climatic class 4, VDI/VDE 3540 Bl.2 Allowed relative humidity Yearly average < 75 % 30 days/year: 95 % other days:...
Page 44
400 ... 1500 l/h pump Operating and display Permeate pump 5-step rotary switch elements Power supply Toggle switch Permeate pump Quartz-controlled WTW peristaltic pump with 4 pump channels. Material V4A steel Test certificates Instrument protection Protective class I according to IEC 1010-1/EN 61010-1...
810 001 standing assembly PurCon Bypass Bypass system for precise BP/PurCon 810 007 setting of the feed flow PurCon air outlet Optional air outlet device to EN/PURCON 810 006 improve the retentate runoff Dilution set Set for the dilution of the...
Page 46
Accessories and consumable materials PurCon Detergent concentrates – Household hygienic cleaner on a sodium hypochlorite basis, e.g. Dan Klorix from the Colgate-Palmolive GmbH company, available in household shops, household departments of department stores, DIY stores, etc. – Trade name: 9062 sodium hypochlorite solution Item no.: 90620.3...
PurCon Index Index Air filter Mains socket Air supply Maintenance activities Amount of permeate supply Mechanical data Authorized use Name plate Clamp strap 75, 78 Cleaning procedure Operating and display ele- Connection data of feed ments pump Operating elements Connections...
Page 48
Index PurCon Taking out of operation Terminals Test certificates Transport damage Valve settings Valves Wall mounting Warning instructions and symbols...