10.8 Procedures
(CH1): Frequency
(measure/source)
[EN] English - K0460
Table 10-7: Voltage (source) error limits
Source
Calibrator
V
uncertainty (V)
0
0.000004
3
0.000010
6
0.000018
9
0.000027
12
0.000036
It is only necessary to do one frequency calibration; use the
measure function or the source function.
Frequency calibration (measure function): With the
necessary calibration equipment, use this procedure:
1. Connect the applicable calibration equipment in
for example: see
Section 3.2.8 (Example procedure:
Measure or source frequency
2. Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
3. Set-up the equipment with these conditions:
Signal generator: Output = 10V, unipolar,
DPI 620-IS:
4. Use the calibration menu
calibration then do the calibration checks
Frequency calibration (source function): With the
necessary calibration equipment, use this procedure:
1. Connect the applicable calibration equipment in
for example: see
Section 3.2.8 (Example procedure:
Measure or source frequency
2. Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
Permitted
DPI 620-IS error (V)
0.00042
0.0006
0.00078
0.00096
0.0011
signals), example A.
square wave
Frequency = 990 Hz
Input units = Hz
Input trigger level = 5 V
(Section
10.3) to do a one-point
(step
signals), example B.
Calibration procedures 10-7
Issue 1
Table
10-1;
5).
Table
10-1;