GE Druck DPI 620-IS User Manual page 86

Advanced modular calibrator
Table of Contents

Advertisement

Issue 1
Calibration check
10-8 Calibration procedures
3. Set-up the equipment with these conditions:
Frequency meter: Gate time = one second
DPI 620-IS:
4. Use the calibration menu
calibration then do the calibration checks
5. To make sure the frequency calibration is correct, set-up the
equipment to do one of these calibration checks:
Frequency (measure) calibration check
Signal generator: Output = 10 V, unipolar,
DPI 620-IS:
Frequency (source) calibration check
Frequency meter: Gate time = one second
DPI 620-IS:
6. Measure or source the specified values
Table
10-9): Hz then kHz. Make sure the error is in the
specified limits.
Table 10-8: Hz error limits (measure/source)
Measure/
Calibrator
source Hz
uncertainty (Hz)
100
0.0002
990
0.0005
5000
0.0002
Waveform = Square;
Amplitude = 10 V;
Frequency = 990 Hz
(Section
10.3) to do a one-point
(Section
square wave
Input trigger level = 5 V
Units: Hz or kHz as specified in
Table 10-8
or
Table
10-9.
(Section
Units: Hz or kHz as specified in
Table 10-8
or
Table
10-9.
(Table 10-8
Permitted
DPI 620-IS error (Hz)
(measure)
0.0023
0.0050
0.00035
(step
5).
3.2.8):
3.2.8):
or
(source)
0.0026
0.0053
0.000185