10-4 Calibration procedures
3. Use the calibration menu
calibration (Zero and FS):
CH1 (one range): 24 mA.
CH2 (two ranges): -24 mA and 24 mA.
4. To make sure the calibration is correct, select the applicable
Current (source) function; see
and apply these values:
CH1/CH2 (24 mA range): 0.2, 6, 12, 18, 24.
CH2 (-24 mA range): -0.2, -6, -12, -18, -24.
5. Make sure the error is in the specified limits
Table 10-3: Current (source) error limits
Source
Calibrator
mA
uncertainty (mA)
±0.2
0.00008
±6
0.00023
±12
0.00044
±18
0.0065
±24
0.0012
(Section
10.3) to do a two-point
Section 3.2.2
or
Section 3.2.6
(Table
Permitted
DPI 620-IS error (mA)
0.0010
0.0016
0.0022
0.0028
0.0034
K0460 - [EN] English
10-3).