11.1 Introduction
Display
Operating temperature
Storage
temperature
Ingress Protection
Humidity
Shock/Vibration
EMC
Electrical safety
Pressure safety
Approved
Battery power
[EN] English - K0460
Chapter 11: General specification
For a full specification of the DPI 620-IS calibrator and its
related accessories (MC 620-IS carrier, PM 620-IS module,
and PV 62x-IS pressure stations) refer to the datasheet
supplied on the CD (CD: P/N UD-0003).
Table 11-1: General specification
OLED: Colour display with touch-screen
-10 to 40°C (14 to 104°F)
-20 to 70°C (-4 to 158°F)
IP65 (DPI 620-IS calibrator only)
0 to 90% relative humidity (RH) non-condensing
Def Stan 66-31, 8.4 cat III
Electromagnetic compatibility: BS EN 61326-1:2006
Electrical - BS EN 61010:2001
Pressure Equipment Directive - Class: Sound Engineering Practice (SEP)
CE Marked
NiMH battery (GE Part number: IO620IS-Battery)
Capacity: 4000 mAh; Nominal voltage: 3.6 V
Charge temperature: 0 to 40°C (32 to 104°F)
If the charger senses the temperature is outside this range, it stops
charging.
Discharge temperature: -10 to 50°C (14 to 122°F)
Charge/discharge cycles: > 500 > 70% capacity