mA
Sensor Trim
®
8-6 HART
device operations
A
Type the Analogue value of Primary Variable mA
n
current (actual calibrated value) into the box
a
using the key-pad and then press Set. Repeat
l
this process with 20 mA selected. This will
o
accurately calibrate the transmitter's output 4
g
to 20 mA loop current.
t
r
i
m
Using the DPI 620-IS, a sensor trim on most Primary Variables
can be performed without having to connect any external
reference meters. If the Primary Variable is a pressure device
use the DPI 620-IS and PV 62x-IS pressure generation stations
to provide the pressure source. Alternatively, use DPI 620-IS
and MC 620-IS to accurately measure pressure.
Navigate to the calibration folder and select the sensor trim
method. Follow the on-screen instructions.
In this example a lower trim has been selected.
The DPI 620-IS is set-up to source a Primary
Variable to the transmitter. CH1 has been
configured to output 0°C. This can be done by
selecting the Switch App button at the bottom
of the method window and configuring the DPI
620-IS to simulate an RTD at 0°C. The Analogue
Value of the Primary Variable can be seen by
selecting CH1 on the drop-down menu.
Type the Analogue Value of the Primary Variable
(actual calibrated value) into the box using the
key-pad and then press Set. Repeat this
process with an upper trim value. This will
accurately calibrate the transmitter's sensor
output.