8.3.3 Communicator
attached to a
network
No functions
on channel 2 (CH2)
®
HART
Function is set
250Ω resistor is set to Off
1
b
a
TAP
[EN] English - K0460
Turn on the internal HART
pressing
icon.
6. Complete the electrical connections and continue with the
®
HART
menu operation.
In this example, the calibrator connects directly to a network.
There must be a 250Ω resistor in series with the loop power supply
®
and the HART
device.
1. Set the function for HART
®
Set HART
User Manual
Section 2.8.4
c
®
resistor. This can be selected by
®
:
d
Home display
®
HART
device operations 8-9