2.9.3 Set the Observed settings (source
[EN] English - K0460
Description
Process: Percent Step;
Values to set: Low, High, Step Size (%FS), Dwell, Auto Repeat (On/Off).
Use these buttons to change the value manually by the
specified Step Size.
Use the start and stop buttons to change the value
automatically by the specified Step Size.
Dwell sets the period between each change.
Description
Process: Defined Step;
Values to set: Low, High, Step Size, Dwell, Auto Repeat (On/Off).
Use these buttons to change the value manually by the
specified Step Size.
Use the start and stop buttons to change the value
automatically by the specified Step Size.
Dwell sets the period between each change.
Description
Process: Ramp;
Values to set: Low, High, Travel, Dwell, Auto Repeat (On/Off).
Use the start and stop buttons for this process. Travel
sets the duration of the Ramp.
Dwell sets the period between each change in direction.
)
The Observed function is available as a Source on the task
selections: CH1, CH2, P1, P2. To set the Observed function,
see
Section
2.8.4.
Issue 1
Prepare the instrument 2-17