7.1 Introduction
2
1
7.2 Analysis
Two Tasks
2
TAP
1
Set one channel as the Input (the DUT) and the other as the Reference.
Set the options A to E.
[EN] English - K0460
Chapter 7: Documenting functions
This section gives examples of the documenting functions
available with the DPI 620-IS calibrator. There are two options:
Analysis
(Section
two Channels on the DPI 620-IS calibrator: the device under
test (DUT) and a reference instrument.
It calculates the % span or % reading error for the device and
then gives a pass or fail indication.
Run Procedure
(Section
Calibration software (Optional item) to download a calibration
procedure.
Before starting:
•
Read and understand the "Safety" section.
•
Do not use a damaged instrument.
Note: Use only original parts supplied by the manufacturer.
To use the Analysis function, set the functions in the two
channels for comparison
procedures:
Documenting
Analysis
3
A
6
Reference
7.2): This function compares the data from
7.3): This uses the Intecal
(Section
2.8.4). Then use these
4
5
7
Input
8
Documenting functions 7-1
A
B
C
D
E
Continued