Table 3: Permitted 1 and 2 function operations
Function
T1
(1)
mA
(1) x
mA(24V)
(1) x
x
x
x
x
IDOS
(1) (2)
= DPI 822 only

To start - Set up the settings

When the task is set up (Table 2/3), use the Settings menu
to adjust the input and/or output operation.
Display: Task
Settings selection
1
T1 + mA
Sk1 = Start/Stop
Sk2 = Settings
Menu:
4A
Settings
(Table 4)
Menu:
4B
Settings
(Table 5)
If there is additional data for a menu option, select
Settings (
) to see the values that are set up. If
necessary, adjust the values. Refer to "Edit functions".
4 - [EN] English
T1
mV
mV
(1)
(1)
(1)
(2)
x
(2)
(2)
x
(2)
(2)
x
(2)
x
x
x
(2)
(2)
(2)
2
3
(If applicable)
Settings
5A
6A
x:y
[ ]/[ ]
Sk1 = Edit
5B
[ ]/[ ]
Table 4: Menu options - Settings (Input)
Options
IDOS
(If applicable)
(1)
... Units To select the temperature units (°C or °F).
(2)
(2)
(2)
... type To select an applicable thermocouple
(2)
x
Description
UPM only = "Pressure Units" if you select an IDOS
task (Table 2/3). Select one of the fixed units of
measurement (psi, mbar ... ).
type (K, J, T ... )
To include maximum, minimum and average
values for the measurement task.
To select and set up a tare value for the
measurement task (a specified value or the reading
on the display).
The instrument subtracts a positive tare value, and
adds a negative tare value.
Additional data: Select Settings (
To select and set up a scale of values: One local
scale for each measurement task (Maximum: 5).
Additional data (Example 1/2): Select Settings (
To select and set up the filter values to give a
smoother output for the measurement task:
Band as a % of full scale (FS). The filter
compares each new value with the previous
value. If the new value is outside the band, it is not
filtered.
Low pass filter time constant in seconds.
Increase the value to increase damping
factor.
Additional data: Select Settings (
To select and set up the alarm values for the
measurement task (maximum and minimum).
Additional data: Select Settings (
UPM only. Gage sensors or sensors with differential
operation. A zero correction that makes the
instrument read zero at local pressure.
Leak Test only. To set an applicable period for the
leak test (Hours:Minutes:Seconds).
)
)
)
)
K346 Issue 2