Table 3: Menu options - Task selections
(Dual Function, area
Options
Description
(If applicable)
White button = A Dual Function is set.
Black button = Dual Function, area
Input measurement tasks:
mA Measure ±55 mA
V Measure ±30V
mA/24V Measure ±55 mA (24V loop power is on)
A switch test
Only when an IDOS UMM is attached. An IDOS
measurement task.

To start - Set up the settings

When the task is set up (Table 2/3), use the Settings menu
to adjust the input and/or output operation.
Display: Task
Settings selection
1
Hz + mA
Sk1 = Start/Stop
Sk2 = Settings
Menu:
4A
Settings
... Units
(Table 4)
Menu:
4B
Settings
(Table 5)
If there is additional data for a menu option, select
Settings (
) to see the values that are set up. If
necessary, adjust the values.
4
- [EN] English
)
is set to off.
2
3
(If applicable)
Display:
5A
6A
... units
bar
Pa
(Table 4)
5B
[ ]
Table 4: Menu options - Settings (Input)
Options
Description
(If applicable)
... Units Pressure Units (UPM only). If you select an IDOS task
(Table 2/3). Select one of the fixed units of
measurement (psi, mbar ... ).
Temperature Units (RTD or TC only). To select the
temperature units (°C or °F).
Frequency Units (Hz only). To select one of these
units:
Hz: Range < 1000Hz
counts/minute (cpm)
...
(TC only). Change the measurement operation:
Temperature to mV OR
mV to Temperature
CJ ... (TC only). To select the type of cold junction (CJ)
compensation.
Automatic: The instrument monitors the CJ
temperature and applies the necessary CJ
compensation.
Manual: Measure the CJ temperature and set the
applicable value. The instrument uses this value to
apply the necessary CJ compensation.
... type Select RTD Type (RTD only). To select an applicable
RTD type (Pt50, Pt100 ... )
Select TC Type (TC only). To select an applicable
thermocouple type (K, J, T ... )
7A
Trigger level
(Hz only). To set the amplitude at which the
instrument senses a frequency signal. Default = 5V.
Auto Detect [ ]/[ ]: Set this option to make the
instrument calculate the value from the available
signal.
(UPM only). Gage sensors or sensors with
differential operation. A zero correction that makes
[ ]
the instrument read zero at local pressure.
(Leak Test only). To set an applicable period for the
leak test (Hours:Minutes:Seconds).
Table 5: (Part of table) Menu options - Settings (Output)
Options
Description
(If applicable)
... Units Pressure/Temperature: Refer to Table 4.
Frequency Units (Hz only). To select one of these
units:
Hz: Range < 1000Hz
pulses/minute (ppm)
...
(TC only). Change the output operation:
Temperature to mV OR
mV to Temperature
CJ ... (TC only). Refer to Table 4.
... type Refer to Table 4.
Amplitude
(Hz only). To set the amplitude of the output signal.
Amplitude = 5V (Default).
To select and set up a value for the "Nudge" output.
Example: 1.000 mA increments.
Additional data: Select Settings (
kHz: Range 0 to 50kHz
counts/hour (cph)
kHz: Range 0 to 50kHz
pulses/hour (pph)
)
K405 Issue 1