Keithley 194 Instruction Manual page 147

High speed voltmeter
Table of Contents

Advertisement

PERFORMANCE
VERIFICATION
5.5 TEST EQUIPMENT
CONNECTIONS
Figure 5-l shows the basic scheme for connecting
the DC
calibrator to the Model 194. While channel 1 connections
are shown, connections
for channel 2 on those units so
equipped would be similar. When making the connection,
use only good quality shielded cable. One end of this cable
should be equipped with a BNC connector to mate with
the VOLTAGE INPUT jack on the instrument,
while the
other end of the connector can be terminated
with a dual
banana jack. When making the calibrator connection,
the
cable shield should be connected to the low output
ter-
minal of the DC calibrator.
CAUTION
Maximum
input
is 250V
peak,
2 x 107 V-Hz.
Figure
5-l.
DC Verification
Connections
5.6 VERIFICATION
PROCEDURE
The steps below outline the procedure for verifying
instru-
ment accuracy. This procedure is intended for use only by
qualified personnel using accurate and reliable test equip-
ment. If the instrument
is out of specifications, refer to Sec-
tion 7 for calibration
procedures, unless the instrument
is
still under warranty. Figure 5-2 shows a flow chart of the
verification
procedure.
Figure
5-2.
Flow
Chart
5-2