Table 7-15. A/D Converter
Checks (Cont.)
step
Item/Component
34
U3?l, pin 12
35
U37I, pin 16
36
Range
37
VOLTAGE INPUT jack
38
Tl'l
(Near A/D/11)
39
DC calibrator
40
TPl
41
Display
42
U376, pins 3-6, 11.14
43
U385, pins 3-6, 11-14
44
DC calibrator
45
Display
46
U376, pins 3-6, 11-14
47
U385, pins 3-6, 11-14
Required
Condition
Remarks
Data burst pulses every 10rsec.
Odd serial data.
500nsec negative-going
pulse
A/D trigger pulse.
every 10psec during
measurement.
Select 3.2V range.
Apply
+3.2767V
Use DC calibrator.
+5.oov
Sample/hold
input
Select -3.2768V output.
-full
range input.
-5.oov
Sample/hold
input
Adjust for -3.2768V on 194
display.
All pulse low during
Odd bit A/D data with
-full
measurement.
range input.
All pulse low during
Even bit AID data with
+full
measurement.
range input.
Select +3.2767V output.
+full range input.
Adjust calibrator for +3.2767V
on 194 display.
All pulse high during
Odd bit data with
+full range
measurement.
input.
All pulse high during
Even bit data with
+full range
measurement.
input.
7-36