Keithley 194 Instruction Manual page 42

High speed voltmeter
Table of Contents

Advertisement

OPERATION
The Uprange key is also used with many other front panel
operations
such as data entry and various XY modes.
These aspects are covered in the respective paragraphs.
3.3.3 Downrange
The Downrange
key operates much like the Uprange key,
except, of course, for the fact that the instrument
is moved
downrange. Once the lowest range is reached, this key has
no further
effect. Pressing downrange,
will also cancel
autoranging,
if that mode was previously
selected.
Like the Uprange
key, Downrange
is used with
other
modes including
data entry and various XY modes.
3.3.4 Range Selection
Considerations
Generally, the lowest range that can be used without
over-
ranging the instrument
is the best one for most situations
(the instrument
will display the "OFLO"
error message
if an overrange signal is applied). Doing so will generally
result in the best overall accuracy. However, because the
instrument
operates a little differently
than an ordinary
DMM, there are some key points to keep in mind when
selecting a range.
When the instrument
is measuring a signal, it is actually
taking a series of samples at pre-programmed
intervals.
As these samples are taken, the resulting
data arc stored
in memory
as a complete measurement.
If the signal
amplitude
varies with time, it is possible that one or more
of these samples is an overrange value. Under these con-
ditions, the unit will still display the "OFLO" message even
if most samples are on range. If this condition
exists, it is
still possible to go back and access the good samples by
using the RECALL button.
When the instrument
is in the continuous
trigger and
autorange mode, it continuously
samples the signal at the
selected rate and attempts to choose the best range based
on the maximum buffer sample amplitude. However, there
could be some situations
where an occasional overrange
reading occurs because the instrument was unable to move
uprange fast enough to keep up with the applied signal.
While autoranging,
the instrument
will take a complete
measurement and, if necessary, change range; it will repeat
measurements
until a satisfactory range is found.
3.4 DATA ENTRY
Many front panel modes such as zero value, delay, trigger
level, rate, and samples require that numeric
data bc
entered from the front panel. Model 194 data is cntrred
by using the ENTER, CANCEL,
FREQITIME,
0 through
9, +/-,
., MHz, kHz, and IIz keys. The operation of these
kevs or kev uoutx
is discussed
in the
follm\Gg
p&graphs.'
"
'
DATA ENTRY KEYS
3.4.1 Data Keys
The data keys, 0 through
9, +/-,
., MHz (,,s), kHz (ms,
mV), and Hz (s,V) are used to enter a numeric quantity
into those modes requiring
data entry. These keys are
operational only when the instrument
has been placed into
the data entry mode by pressing another mode button re-
quiring data entry. For example, pressing RATF will enter
this mode and allow you to program a rate.
During the data entry process, a flashing digit or segment
(the cursor) on the display will indicate which digit \vill
be affected by a numeric key press. The 0 through 9, ., or
+/-- keys can be pressed at the appropriate
times to enter
the desired data. During the numeric entry pnxcss,
the
cursor can be moved to the right or left by pressing
Uprange or Downrange
respectively,
and '1 single-digit
change can be made at that point.
3-3

Advertisement

Table of Contents
loading

Table of Contents