Keithley 194 Instruction Manual page 47

High speed voltmeter
Table of Contents

Advertisement

OPERATION
2. Press: SAMPLES,
FREQiTIME
(if necessary to enter
data as time duration),
6, 0, 2, ms.
3. Note that the 602msec programmed
value has chang-
ed to 600msec, which is the nearest integer multiple
of
the l2msec sampling
interval.
Example 4: Setting the sampling
duration
smaller than
the sampling
interval.
The sampling
duration
cannot be smaller than the sam-
pling interval, as discussed previously. The instrument will
round off the value if you attempt to program the instru-
ment in this manner, as in the example below.
1. Press RATE, FREQiTIME
(if necessary to place the
display in the time interval mode), 1, 0, 0, ms, ENTER.
This keystroke sequence programs a 10011~ sampling
interval.
2. Press SAMPLES, FREQiTIME
(if necessary to display
and enter time units), 2, 0, ms.
3. Note that the measurement duration is changed to Opsec
(1 sample) because it is rounded
down.
Example 5: Automatic recalculation
of number of samples.
If the sampling
rate is changed
after programming
measurement
duration,
the instrument
will automatical-
ly recalculate the number of samples to keep the measure-
ment duration
the same, as in the example below.
1. Press RATE, FREQ/TIME
(if necessary to enter time in-
terval information),
1, ms. At this point a lmsec sampl-
ing interval has been programmed.
2. Press SAMPLES,
FREQiTIME
(if necessary to enter
duration as time information),
5, 0, 0, ms, ENTER. This
sequence enters a 500msec sampling
duration.
3. Press SAMPLES and FREQiTIMlZ to display the number
of samples. The display should indicate that the pro-
grammed number of samples is 501 because the sampl-
ing duration
is 500msec, and the sampling
interval is
1lllSeC.
4. Change the sampling interval to 5msec as follows.Press:
RATE, 5, ms.
5. Press SAMPLES and FREQiTIME
to display the pro-
grammed number of samples. Note that the number
of samples has been changed to 101 because of the
change in sampling
interval,
although
the sampling
duration remains at 500msec, as previously
programm-
ed (you can verify
these values
by pressing
the
FREQiTlME
key while in the samples mode).
3.5.3 Samples
and Rate Selection
Considerations
Because the Model 194 can sample input signals at rates
as high as lMHz,
it is ideal for many applications
involv-
ing the analysis of many time-varying
signals, both of the
periodic and transient variety. To ensure optimum accuracy
when measuring
such signals, care must be taken when
selecting
both the sampling
rate and the number
of
samples. In the following paragraphs, we will discuss some
of these considerations
that should be taken into account
when
choosing
sampling
rates
and
measurement
durations.
Input signals to the Model 194 are in analog form. Inter-
nally, however, the Model 194 operates in the digital world.
Thus, the analog signal must be converted into digital in-
formation
by the A/D (Analog-to-Digital)
converter of the
instrument.
This conversion process is not continuous,
but
rather is done at discrete intervals,
determined
by the
entered rate parameter. If the sampling
rate is too slow,
considerable information
about the original analog signal
will be lost, and errors can creep into the resulting
data.
For example, assume that a sinusoidal
waveform is being
sampled at regular intervals, and the result digitized and
stored internally.
Once the data is sampled and stored, we can attempt to
reconstruct the original waveform from the data. However,
the result is no longer a smooth, continuous
waveform,
but is instead made up of discrete steps. Thus, as the result
of this digitization
process, we may have lost much im-
portant information
about the original
signal.
To make the steps in the reconstructed
data smaller, we
can increase the sampling rate. At the same time, we will
have to increase the number of samples per measurement
if we still wish to measure at least one cycle of the applied
signal.
It is clear then, that we should make the sampling rate suf-
ficiently high so as not to loose important information
pre-
sent in the original input signal. Information
theory states
that, for sinusoidal
waveforms,
the sampling
frequency
must be at least twice as high as the highest frequency
component
in the measured sienal. Thus. if a 1OOkHz
signal is to be sampled, the sampling frequency must be
at least 200kHz.
3-8

Advertisement

Table of Contents
loading

Table of Contents