IEEE-488 PROGRAMMING
Model 8573 Programming Example-Alter
instrument
con-
figuration
from the factory default value with several front
panel controls and enter the following
statement into the
IBM PC keyboard:
When the return key is pressed, the instrument
returns
to default conditions.
4.8.8 SDC (Selective
Device Clear)
SDC is an addressed command that performs essentially
the same function
as the DCL command. However, since
each device must be individually
addressed, SDC provides
a method for clearing only a single, selected instrument
instead of clearing all devices simultaneously,
as is the case
with DCL. When the Model 194 receives the SDC com-
mand, it will return to the default configuration.
The fac-
tory default configuration
is shown in Table 4-7 but many
modes except for the primary address may be different if
you change the SETUP 1 (Al) configuration.
To transmit the SDC command,
the controller
must per-
form the following
steps:
1. Set ATN true.
2. Address the Model 194 to listen.
3. Place the SDC command byte on the data bus.
HP-85 Programming
Example-Using
several front panel
controls, alter instrument
modes from the factory default
configuration.
Send SDC with the following
statement:
When the above statement is executed, the instrument
returns to the default configuration.
Model 8573 Programming Example--Alter
instrument con-
figuration
with several front panel controls and enter the
following
statements into the IBM PC:
When the return key is pressed, the instrument
returns
to default configuration.
4.8.7 GET (Group Execute
Trigger)
GET may be used to initiate a Model 194 measurement se-
quence if the instrument
is placed in the appropriate
trig-
ger mode (see paragraph 4.9). Once triggered, the instru-
ment will perform the measurement
sequence in accor-
dance
with
previously
selected
rate
and
sample
parameters.
To send GET, the controller
must perform the following
sequence:
1. Set ATN low.
2. Address the Model 194 to listen.
3. Place the GET command byte on the data bus.
HP-85 Programming
Example-Type
in the following
statements to place the instrument
in the correct trigger
mode for purposes of this demonstration:
,?EI',OTE ;:'iij's
Now trigger the measurement
sequence by sending GET
with the following
statement:
"r F: 1: 1: 1; E h: ;7 $3'3
When the END LINE key is pressed, the measurement se-
quence will be triggered.
Model 8573 Programming
Example-Type
in the follow-
ing statements to place the instrument
in the correct trig-
ger mode for this demonstration:
Trigger the measurement
sequence by sending GET with
the following
statement:
1: $4,. ,... 1 R ,- F: 1; 8:: p, 1, '3 ,:* :.; ::I
The measurement
sequence will be triggered when the
above statement is executed.
4-14