Basic Sampling - Keithley 194 Instruction Manual

High speed voltmeter
Table of Contents

Advertisement

GETTING STARTED
Measurement-a
series of samples stored in internal
memory. You can control the number of samples and how
fast they are taken with the SAMPLES and RATE keys.
Reading-a
measurement that is mathematically
processed
in some way and then displayed
on the front panel
OI
transmitted
over the IEEE-488 bus. Typical processes in-
clude waveform (display a single sample), average (average
the samples in the measurement
and display the result),
and standard deviation (take the standard deviation of the
measurement
and display it).
2.6.2 Sampling
Discussion
Using the procedure discussed in paragraph 2.5, the Model
194 appears to operate much like an ordinary
DMM,
in
that a reading immediately
appears on the display. Actual-
ly, the instrument
is taking a number of samples, digitiz-
ing the analog value, and storing the result in its internal
memory. The resulting
sequence of samples is called a
measurement.
While in the continuous
trigger mode (as
in this example), the display is continuously
updated with
the sample stored in the memory location at the trigger
point (assuming the waveform mode is in effect). Samples
stored in the remaining
locations can be accessed by us-
ing the RECALL button.
Figure 2-9 demonstrates
the basic idea behind
signal
sampling. Here, a time-varying
signal with the amplitude
shown is being sampled at specific intervals. As each sam-
ple is taken, it is digitized into an S-bit or 16.bit binary value
(depending
on the sampling rate and number of samples
in the measurement) and stored in memory. The complete
sampling sequence is a measurement,
as defined above.
Once the measurement sequence is performed,
a reading
can be generated by processing the block of samples in
s"me way. For example, to obtain the true RMS value of
the measurement
(remember a measurement
is a series
of samples), you would use the TRMS function.
By using
the instrument
in a single trigger mode, you could apply
a variety of different
mathematical
functions
to a single
measurement.
Conversely, a single mathematical
process
could be applied to a variety of different signals by using
the instrument
in the continuous
trigger mode. A single
sample can be displayed by using the waveform mode (in
which case the sample at the trigger point is displayed),
or by using the RECALL button
to display
individual
SCl"lpleS.
v
Figure 2-9. Basic Sampling
2.7 TYPICAL
OPERATING
MODES
The following
information
will help you t" select the
various
operating
modes for commonly
encountered
voltages and waveforms. Keep in mind that these are in-
tended only as a starting point. Some experimentation
may
be required
to determine
the optimum
instrument
con-
figuration
for a particular
measuring
situation
based on
your particular
waveform analysis requirements.
Table 2-5 summarizes
recommended
mode selection for
four c"mm"n
voltages or waveforms: a DC voltage of 3Ov;
a 60Hz, 2.82V RMS sine wave; a ZV, lkHz symmetrical rec-
tangular waveform;
and a 50V peak, 15.734kHz sawtooth
waveform.
When selecting operating
modes, the following
points
should be kept in mind:
1. The selected range should be high enough to handle
the peak value of the waveform you are measuring. For
example, the peak value of a 30V RMS sine wave is
42.4V. Thus, you would have to place the instrument
on the ZOOV range to properly measure this waveform.
Keep in mind that the instrument
will normally display
the OFLO (overrange) err"r message even if only one
sample is overrange.
2-16

Advertisement

Table of Contents
loading

Table of Contents