Displaying and Setting ω
This machine is provided with a function to perform an integrator (ω
2
ω
T operation, set the ω
Step
1
2
3
4
Thermo Scientific
2
T
2
T value instead of the run time.
Key operation
Screen display and notices
Set the speed in the
See Basic Operation.
Run Screen.
The FUNCTION indicator turns to the ω
1.
Use cursor keys
to move cursor
to ω
2
T on the
RUN SCREEN.
2.
Press the ENTER
key.
2
Enter a desired ω
Entered value is displayed on the ω
T
value.
The TIME display shows "- : -".
Example: 9.87E11
Set ω
Computed and integrated value is displayed on the ω
When you press
START key, the run is
controlled by the set
2
ω
T value rather than
by time.
2
T SET column and the message indicator respectively.
2
T

Displaying and Setting w2T

2
T) operation. To perform an
2
T set screen.
2
T indicator.