Figure 4-8. Operating Mode (RTC)
2. Note that the RTC setting is not available in the following cases:
3. Clear the RTC setting and then reset the run time if it is necessary to change the run time setting
after setting the RTC operation.
4. Recall the saved run conditions and enter the RTC setting when performing the programmed
operation (including the step-mode operation) and the RTC operation in combination. The
centrifuge automatically computes the total run time of all steps in the programmed operation and
also the start time for RTC operation. Note that the saved run conditions cannot be recalled after
setting the RTC operation.
5. Press the STOP key to stop the operation. The RTC operation is stopped and the rotor stops.
6. When performing the RCF operation and the RTC operation in combination, enter the RCF
setting first then enter the RTC setting. (See Displaying and Setting RCF.)
7. When performing the ω
setting first then enter the RTC setting. (See Displaying and Setting

Displaying and Setting RCF

The RC6 Plus Superspeed centrifuge, has in its internal memory, data representing the maximum radii
of all available rotors. Based on this data, the centrifuge automatically computes relative centrifugal
force (RCF) values from set speed, or speed from set RCF values, and then displays the result on the
control panel. This section explains how to use this RCF displaying and setting capability of the
centrifuge.
Thermo Scientific
1
2
i.
The time setting on the RUN SCREEN is "HOLD" (continuous run).
Change the run time setting from "HOLD" to a desired numeric value.
ii. The start time has already passed.
Change the setting so that the start time will be later than the current time.
iii. The start time is 20 days or more from the current time.
Change the setting so that the start time will be within 20 days.
2
T operation and the RTC operation in combination, enter the ω
4
3
4
OPERATION
Displaying and Setting RCF
5
1
ω
2
T.)