Step-Mode Operation - Thermo Scientific Sorvall RC6 Plus Instruction Manual

Superspeed centrifuge
Hide thumbs Also See for Sorvall RC6 Plus:
Table of Contents

Advertisement

Step-mode Operation

This centrifuge has the step-mode operation capability that allows you to save three different sets of
values for a run parameter set in a single memory location (MEMORY Nos. 31 - 33, 41 - 43, and 51 -
53) and then change some or all of the run conditions (e.g., speed, run time, rotor temperature, etc.)
for each step during a step-mode run. Save step-mode run conditions at the MEMORY Nos. 31 - 33
(41 - 43 or 51- 53) in accordance with "3-3-1 Programmed Operation (1)".
When the MEMORY No. 31 is recalled, the centrifuge automatically performs step-mode operation in
order of MEMORY Nos. 31, 32 and 33.
Procedure for step-mode operation
Example
The table below shows the run parameters and their values required for an example 3-step run
(MEMORY Nos. 31 to 33). Fig. 4-6 depicts how the example run will proceed.
Speed
Run time
Temperature
Accel rate
Decel rate
Rotor No.
Figure 4-6. Details of an example step-mode operation
Note
1. Select the same rotor number for each step. Otherwise, alarm message "ROTOR NO." is
indicated and the centrifuge stops operation. (Contents of the memory cannot be called up.)
2. Use the MEMORY Nos. 32 and 33 (42 and 43 or 52 and 53) for the step-mode operation with
two steps.
Thermo Scientific
1st step
2nd step
(Memory No. 31)
(Memory No. 32)
1000 rpm
20000 rpm
30 min
60 min
4°C
4°C
9
9
9
9
05
05
3rd step
(Memory No. 33)
5000 rpm
10 min
4°C
9
7
05
Sorvall RC6 Plus
4
OPERATION
Step-mode Operation
4-13

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents