Centrifuge Specifications - Thermo Scientific Sorvall RC6 Plus Instruction Manual

Superspeed centrifuge
Hide thumbs Also See for Sorvall RC6 Plus:
Table of Contents

Advertisement

1
INTRODUCTION & DESCRIPTION

Centrifuge Specifications

Centrifuge Specifications
Table 1-1. Centrifuge Specifications
Run Speed
Maximum Relative Centrifugal Force
Run Temperature
Run Time Selection Range
Ambient Temperature Range
Mass (Weight)
Maximum Capacity
Dimensions
Noise Level
Heat Output
*
Speed in revolutions per minute (rpm) is related to angular velocity, , according to the following:
=
rpm
Where = rad/s. All further references in this manual to speed will be designated as rpm.
**
The centrifuge will operate at ambient temperatures up to 40°C, but refrigeration system performance may be less than optimal above
25°C.
After the centrifuge system has reached equilibrium. Ambient air temperature at the centrifuge air inlets must be between 2°C to 40°C
with maximum relative humidity of 80% at 31°C, linearly decreasing to 50% at 40°C. If the ambient air temperature is above 25°C, the
centrifuge may not maintain low temperatures at high speeds, therefore, avoid areas near heat sources (for example, direct sunlight,
heating pipes and radiators). Also, avoid close grouping of centrifuges or other heat-producing laboratory equipment.
***
After the centrifuge system has reached equilibrium. Ambient air temperature at the centrifuge air inlets must be between 2°C to
40°C with maximum relative humidity of 80% at 31°C, linearly decreasing to 50% at 40°C. If the ambient air temperature is above 25°C,
the centrifuge may not maintain low temperatures at high speeds, therefore, avoid areas near heat sources (for example, direct sunlight,
heating pipes and radiators). Also, avoid close grouping of centrifuges or other heat-producing laboratory equipment.
****
As measured with SE-12 rotor at 22,000 rpm (with compressor turned on) 3 feet from the front of the unit and four feet from the floor.
*****
For the SS-34 rotor spinning 20,000 rpm at 4°C, after it has reached equilibrium. Other rotors, speeds, and temperatures cause the
heat output to vary.
1-4
Sorvall RC6 Plus
*
Speed Selection Range (rpm)
Speed Control Accuracy
Temperature Selection Range
Temperature Control Range
Temperature Control Accuracy
Width
Height to top of control console
Height with door open
Depth (including rear duct assembly)
Depth (without rear duct assembly)
From bottom to chamber inlet
2
 0.10472
------ -
=
rpm
60
300 to 22,000
±25 rpm
52,200 x g (F22/Micro)
-20°C to 40°C
2°C to 40°C
±2°C
1 minute to 99 hours, 59 minutes
HOLD for continuous operation
2°C to 40°C
350 kg
4,000 ml (F10S-4x100 LEX)
752 mm
1112 mm
1500 mm
835 mm
685 mm
858 mm
<64 dB
*****
3.7 kW (12,600 BTU/h)
Thermo Scientific
**
***
****