4
OPERATION
Basic Operation
Item
Example set value
1
Press a cursor to turn the display to
ready-to-enter state.
2
Move the cursor to the desired item by
pressing the cursor keys.
(The arrow marks on the cursor keys
indicate cursor directions.)
3
The cursor blinks on the entry line for 30
seconds.
The display is now ready-to-enter state.
4
Enter the desired value with
the ten-key numerical pad.
Entered numerals are shifted to the left in
order.
5
Check the setting and fix it by pressing the
ENTER key.
Setting can also be fixed by pressing a
cursor key. The CE key is used to cancel
the setting.
6
Press the START switch.
7
After the rotor has come to a stop, lift the
door handle. Then remove the rotor.
Setting range
Note Keep the chamber door closed after the rotor has been removed to inhibit the formation of
condensation on the chamber walls.To repeat the run with the same speed, time and temperature
parameters, install the rotor, close the chamber door, press START and then press ENTER.
4-6
Sorvall RC6 Plus
Speed (SPEED)
20,000 rpm
The display turns to readyto-enter state.
The last zero (0) need not
be entered.
The speed setting is "20,000 rpm".
300 rpm to the maximum speed: in
increments of 10 rpm under 10,000 rpm,
and in increments of 100 rpm over 10,000
rpm
Run Time (TIME)
2 hours, 30 minutes
The display turns to ready-to-enter state.
The cursor can be moved to the hours
setting position by pressing the cursor left
key.
For continuous run, press the HOLD key.
The run time setting is "2:30
(2 hours 30 minutes)".
1 minute to 99 hours 59 minutes: in
increments of 1 minute