Leveling The Centrifuge - Thermo Scientific Sorvall RC6 Plus Instruction Manual

Superspeed centrifuge
Hide thumbs Also See for Sorvall RC6 Plus:
Table of Contents

Advertisement

• The unit is intended for indoor use only. Ambient air temperature at the centrifuge air inlets must
be between 2°C to 40°C with maximum relative humidity of 80% at 31°C, linearly decreasing to
50% at 40°C. If the ambient air temperature is above 25°C, the centrifuge may not maintain low
temperatures at high speeds, therefore, avoid areas near heat sources (for example, direct sunlight,
heating pipes and radiators). Also, avoid close grouping of centrifuges or other heat-producing
laboratory equipment.
• Maintain a 30-cm clearance around the centrifuge. Do not store any substances in the clearance
envelope.
Note The RC6 Plus is a Class A product and is not intended for home use. If used in a domestic
environment you must contact your local authorities for authorization.
Figure 2-1. Clearances for the Centrifuge

Leveling the Centrifuge

1. Turn on the centrifuge power and open the door. (If power supply is not yet available, perform the
Emergency Recovery from Power Failure procedure discussed in Chapter 4.)
2. Place the level across the top of the drive spindle.
3. Using a 24 mm wrench, turn each of the four level adjusters until the casters are 5 to 15 mm off
the floor.
4. Level the unit by adjusting the level adjusters.
5. When the centrifuge is level, make sure each of the four level adjusters is secure and rattle-free.
6. Secure each of the four level adjusters with the lock nuts.
Thermo Scientific
CAUTION Leave adequate space for airflow around the centrifuge and make sure no vents
are blocked. Blocking the airflow entering and/or exiting the centrifuge can result in
reduced performance, overheating and possible centrifuge damage.
2
INSTALLATION
Leveling the Centrifuge
Sorvall RC6 Plus
2-3