OPERATION
RTC (Real Time Control) Operation
Step
5
6
7
1. The RUN mode indicator on the panel turns as follows in RTC operation.
4-16
Sorvall RC6 Plus
Key operation
Enter the desired date and
time (month, day, hour and
minutes) using the cursor
keys and the ten-key
numerical pad. Press the
ENTER key.
Select Yes or No in
response to the prompt.
Check the RTC setting on the
RUN SCREEN and press the
START key.
Screen display and notices
• The range for "hour" setting is from 0 to 23 (24-hour display).
• Do not enter any date and time that passed the current time.
• Set a proper stop time considering the centrifugation time so that the start
time will be later than the current time.
• It is not possible to set an operation that will start 20 days or more from
the current time.
When selecting "Yes":
• The screen turns to the RUN SCREEN and the RTC time is displayed.
• RTC appears on the message indicator.
When selecting "No", the display turns to the screen of Step 2
setting again.
• Perform operation according to Operating Procedure.
• Note that the run time setting cannot be changed after setting the RTC
operation. Clear the RTC setting and then reset the run time if necessary.
• The centrifuge turns to "DELAY" mode by pressing the START key and wait
until the set time. The centrifuge automatically starts operation at the set
time and keeps operation during the designated time.
• The RUN mode indicator on the message display turns to "DELAY".