Thermo Scientific Sorvall RC6 Plus Instruction Manual page 27

Superspeed centrifuge
Hide thumbs Also See for Sorvall RC6 Plus:
Table of Contents

Advertisement

Table 3-1.
No
Name
1
Field display
2
FUNCTION field
3
Message
indicator
4
RUN mode
indicator
Thermo Scientific
Display Panel Function (refer to Figure 3-1)
Displays the following run conditions.
For SPEED, TIME and TEMP displays, the upper line shows the actual run state and the
lower line shows the set value. Refer to Setting Run Conditions for details.
• SPEED (Speed display)
(Upper line)
Displays rotor speed in increments of 10 rpm under 10,000 rpm and
increments of 100 rpm over 10,000 rpm.
(Lower line)
Displays rotor speed in increments of 10 rpm under 10,000 rpm and
increments of 100 rpm over 10,000 rpm. Set speed range is from 300 rpm to
22,000 rpm.
• TIME (Run time display)
(Upper line)
Displays remaining run time during operation. If HOLD is selected, displays
elapsed run time.
(Lower line)
Run time range is from 1 minute to 99 hours, 59 minutes in increments of 1
minute.
• TEMPERATURE (Temperature display)
(Upper line)Displays rotor temperature in increments of 1 degree centigrade.
(Lower line)Temperature range is from - 20 degrees centigrade to 40 degrees
centigrade in increments of 1 degree centigrade.
• ACCEL (Acceleration rate display)
Displays 1 to 9 acceleration rates.
• DECEL (Deceleration rate display)
Displays 1 to 9 deceleration rates and free coast (0).
• PROG Used to save run conditions for programmed operation.
• RCF Used to display and set RCF value.
• ω
2
T Used to display and set for integrator operation.
Displays alarm message, prompt and rotor model.
Displays operating mode with illustration of a rotor.
Operating modes are as follows:
STOP, ACCEL, RUN (Displayed when rotor is rotating at set speed), DECEL
And DELAY (Displayed up to RTC operation is started)
3
CONTROLS, DISPLAYS & INDICATORS
Operation Panel
RC 6 Plus
3-3