Important Safety Information

Safety During Installation and/or Maintenance

CAUTION
Do not pour any solution such as water, detergent or disinfectant directly into the rotor
chamber. Otherwise, the bearings of the drive unit may corrode or deteriorate.
Before relocating the centrifuge, remove the rotor from the rotor chamber to avoid damage
to the drive shaft.
Always keep the liquid crystal panel in a visible position while the POWER switch is turned
on, so that you can check the current operating state (running or stopping).
Safety During Installation and/or Maintenance
DANGER
When servicing the centrifuge, be sure to turn off the POWER switch, turn off the
distribution board of your centrifuge room, and then wait for at least three minutes before
removing covers or tables from the centrifuge to avoid electrical shock hazards.
WARNING
Level the centrifuge by using the four level adjusters and secure them completely. Improper
securing can cause significant movement of the centrifuge in the event of a rotor
disengagement.
Installation or relocation of your centrifuge must be done by Thermo Fisher Scientific or a
Thermo Fisher Scientific representative. Contact Thermo Fisher Scientific or your local
representative of Thermo Fisher Scientific products.
Do not position an object so that it is difficult to disconnect the power cord from the outlet.
If you do so, you cannot disconnect the power cord from the outlet when you observe some
abnormality.
CAUTION
Avoid places that are exposed to ultraviolet rays for operation or storage of the centrifuge.
Otherwise, the covers can be discolored and the coating can be peeled off easily. If
installation in such a place is unavoidable, cover the centrifuge with a cloth after operation
to protect from ultraviolet rays.
If you encounter any problem concerning maintenance that is not covered in the operation
manual, please contact Thermo Fisher Scientific for assistance. For addresses, refer to the
rating label on the right side of the centrifuge.