ResMed Elisee 350 Clinical Manual page 77

Table of Contents

Advertisement

Symbol
Message
MINI FIO
MAXI FIO
O
2
WRONG
CIRCUIT
MAXI LEAKS
PATIENT
APNEA
INT BATT
EMPTY
INT BATT
LOW
EXT BATT
EMPTY
EXT BATT
LOW
EXT BATT
MISSING
EXT MISSING
MAINS
MISSING
TEST FAILED
* In pressure-controlled modes, the Pmax alarm threshold cannot be set to a value lower than the delivered Pi + PEEP (or the
delivered pressure support). If the user sets an alarm threshold lower than this pressure value, the alarm threshold setting will
be ignored. In this situation, the delivered Pi + PEEP (or PS + PEEP) becomes the Pmax alarm threshold.
Note: The Battery Flat and Battery Low alarms are deactivated as soon as the Elisée 350 is
connected to the mains or to an external DC power supply. Press
message from the screen.
NOT013281-8
Alarm
Minimum FiO
2
2
Maximum FiO
2
2
FAULT
HP O
missing
2
Device configured for single
circuit (manual test), but a
double circuit is connected
High Leak
Apnea
Internal Battery Flat
Internal Battery Low
External Battery Flat
External Battery Low
No External Battery
External Power Supply
Disconnection
Mains Disconnection
Manual Test Interrupted
Trigger
Priority
threshold [can
be stopped?]
High
FiO
– 10%
2
[NO]
High
FiO
+ 10%
2
[NO]
High
FiO
≠ 21% or
O
2
2
100%
2
min
button pressed and
HP O
≥ 0.5 bar
2
[2 min]
High
V
> 10% of V
TE
TI
[2 min]
High
Alarm setting
[2 min]
Medium
Tapnea
[2 min]
High
Charge < 5%
[NO]
Medium
Charge < 15%
[2 min]
Medium
Charge < 5%
[2 min]
Low
Charge < 15%
[2 min]
Low
-
[YES]
Low
-
[YES]
Low
-
[YES]
High
If one of the four
test steps fails
[2 min]
to remove the alarm
Time to trigger
Maximum of 20 consecutive
cycles below alarm threshold
Maximum of 20 consecutive
cycles above alarm threshold
Immediately
6 cycles
6 consecutive cycles above
alarm threshold
Immediately
Immediately
Immediately
Immediately
Immediately
Immediately
1 s
1 s
-
Alarms and troubleshooting
71/102