WARNING
Set a V
Single circuit
1
2
WARNING
The end of the proximal pressure line must be placed as close to the patient as possible.
Proximal pressure measurement is essential for insufflation pressure monitoring and
pressure trigger sensitivity.

4.3 Oxygen

Installing the oxygen sensor
38/102
alarm threshold based on the measured expiratory volumes.
TE
3
2
15 min.
1. Connect the expiratory valve control tube.
2. Connect the proximal pressure line by pushing the tube onto
the notched connector. Depending on the circuit used, it may
be necessary to cut off the nozzle of the proximal pressure
line.
Patient
3. Connect the inspiratory limb to the insufflation outlet.
Note: Perform a manual test so that the Elisée 350 registers
the new circuit configuration.
1. Switch off the ventilator.
2. Prepare the new oxygen sensor:
Check the expiry date (month and year) on the
sensor packaging. It can be installed at any
time before the date on the label.
Let the sensor sit in ambient air for 15 minutes.
3. On the Elisée 350, remove the protective cap from
the sensor.
3
4. Using the disassembly tool supplied with the
ventilator, place the sensor in its compartment and
screw into place.
Disassembly tool
4
Valve control tube
1
Proximal pressure line
NOT013281-8
2