ResMed Elisee 350 Clinical Manual page 69

Table of Contents

Advertisement

Calibrating the touch screen (while ventilation is stopped)
This operation allows the user to readjust the reactive areas of the touch screen.
TEST
TEST
PRESET
PRESET
SETTING
SETTING
PREVIOUS SETTING
PREVIOUS SETTING
<< Press here
CALIBRATION
BTPS correction
BTPS correction (body conditions: body temperature, ambient pressure and vapour saturated) is
activated by default each time the ventilator is started.
It corrects the inspiratory and expiratory volumes on the double circuit. When the option is
deactivated, the ventilator passes into ATPD mode (Ambient Temperature and Pressure, Dry), the
mode used to calibrate the ventilator.
Note: The BTPS correction varies depending on the atmospheric pressure, the temperature of
the exhaled gases (estimated at 37°C), and the ambient temperature.
RC auto calcul. (invasive ventilation and ACV mode only)
This option, always reset to NO each time the ventilator stops, automatically calculates the
compliance and resistance of the patient's airways every 15 minutes.
The results are displayed on the measurements screen for 5 minutes.
Low-pressure oxygen input
This option activates the low-pressure or high-pressure oxygen supply.
If "Low pressure oxygen input" is set to NO (when the ventilator starts up), the ventilator is
programmed to use high-pressure oxygen.
If "Low pressure oxygen input" is set to YES, the ventilator is programmed to use low-
pressure oxygen.
Selecting target parameters in ACV mode (during ventilation and
while ventilation is stopped)
This parameter is saved in the memory when the ventilator is turned off.
The adjustable values during Assisted Volume-Controlled Ventilation (ACV) are:
The inspiratory time (Ti)
The max. flow (
The I:E ratio.
NOT013281-8
1. On the main screen, press and hold
Last test the 18/06/10
Double circuit : PASS
simultaneously until the calibration screen appears.
NEW
NEW
SETTING
SETTING
ACV
ACV
2. On the calibration screen, press the targets appearing in
various places. Keep pressing until you hear a beep.
The touch screen is now calibrated and the main screen
reappears.
°
max)
and
How to use Elisée 350
63/102