1 Mains power pack
The pack is housed in the compartment located on the underside of the ventilator. It supplies the
device with DC power, converted from mains power, and enables the internal battery to recharge.
Note: If used outside of its compartment, the mains power pack is connected to the ventilator
with an external power cord and is therefore treated as an external power supply (see next
paragraph). Place the dummy pack in the compartment so that it is never left empty.
CAUTION
The mains power pack is not splashproof (IPX0). Therefore, it should not be connected to
the ventilator when there is a risk of water penetration.
2 External DC power supply
An external DC power supply can be provided by:
CAUTION
Only power supplies able to deliver 12
cord to power the ventilator.
External power supply cord
(to connect the mains power
pack to the Elisée)
Note: To switch the ventilator off, disconnect the external DC power supply.
If the ventilator is turned off while connected to an external power supply, it automatically
switches to standby mode to recharge the internal battery (and the external battery, if this is
connected to the ventilator).
WARNING
The external DC power supply should be capable of powering the device for at least one
hour. If the external supply is less than 20 V during ventilation, the batteries will not be
recharged.
10/102
The mains power pack, using the external power cord
A DC power supply connected to the Elisée via the ResMed cigarette lighter connector.
28 V / 15 A can be connected through the DC power
DC power cord with ResMed
cigarette lighter connector for
power supply from vehicle
NOT013281-8