Shutting down the ventilator
20 0
cm
H
O
2
90
80
70
60
50
40
30
20
<AI
10
0
<PE
Each time the ventilator is shut down, an audible signal is heard.
Forced ventilator shutdown
If there is a fault with the touch screen, a forced shutdown may be performed:
1. Disconnect the ventilator from the mains or external power supply.
2. Press
68/102
1. Press the
MENU
MENU
Configuration
Options ...
back
2. Press
Locked touch screen
Pause (stop ventilation)
Stop machine
The message "STOP IN PROGRESS" flashes, then the device
turns off.
3. When the audible alarm sounds, press the
and
at the same time, and hold until the ventilator switches off (10 seconds).
button on the right-hand side of the ventilator.
"Stop machine" for three seconds.
least one second.
button for at
NOT013281-8