2 Checking the patient circuit
Before connecting the patient to the ventilator, check the whole of the patient circuit to eliminate
any risk of leaks or defective parts.
Check that:
The expiratory circuit support matches the circuit type: single or double
The patient circuit and accessories are properly connected to the expiratory circuit support.
Note: Perform a manual test to ensure the ventilator has correctly identified the type of patient
circuit installed.
3 Manual test
This test must always be performed before the ventilator is programmed for a new patient and
whenever the circuit configuration is changed.
WARNINGS
The ventilator must not be connected either to a patient or to an oxygen supply during
this test.
If the test is unsuccessful, check the circuit configuration and repeat the test. Make sure
you complete every step. If the test is again unsuccessful, stop using the ventilator and
contact a technician.
Make sure that when you perform this test, all the accessories that will be used during
ventilation are connected. Repeat the test if an accessory is added or removed.
This quick test (around 1 minute) consists of four stages:
TEST
TEST
48/102
Measurement of the sensor offsets
Checking for internal leaks in the ventilator
Measurement of the inspiratory resistance of the circuit
Measurement of the expiratory resistance and calibrating the expiratory flow (if a double
circuit is connected), the circuit compliance and the O
Las test the 18/06/10
Double circuit : PASS
NEW
NEW
PRESET
PRESET
SETTING
SETTING
SETTING
SETTING
PREVIOUS SETTING
PREVIOUS SETTING
ACV
ACV
Return to main
Last test
screen
le 18/06/10 at 14 : 50 : 35
Back
Circuit test : PASS
Kind of circuit : DOUBLE
Ctot = 0,7 ml/cmH2O
Ri = 0.4 cmH2O (30 l/min)
Ri = 0.7 cmH2O (60 l/min)
Re = 1.4 cmH2O (30 l/min)
Re = 2.9 cmH2O (60 l/min)
RESTART
RESTART
TEST
TEST
Cancel
Unplug oxygen
Unplug patient circuit !
When done, press:
CONTINUE
CONTINUE
sensor gain.
2
1. On the main screen, press the "Test" button.
The results of the last test are displayed.
2. Press the "Restart" button.
3. Follow the instructions displayed (disconnect the
oxygen supply, the patient circuit and its
accessories), then press "Continue". The ventilator
measures the sensor offsets.
NOT013281-8