Obstruct the inspiratory outlet !
When done, press:
CONTINUER
CONTINUER
CONTINUE
CONTINUE
Attach your circuit to the machine
Leave the patient end of the circuit
open.
When done, press:
CONTINUE
CONTINUE
Obstruct the end of the patient
circuit.
When done, press:
CONTINUE
CONTINUE
The results of the test are displayed. They tell you:
Whether the test failed or was successful
What type of patient circuit is connected
The resistance and compliance values of the patient circuit
If the oxygen sensor is defective and must be replaced (if so, a flashing message appears
on the screen). Replace the sensor and repeat the manual test.
7 . If the test was successful, press
to redo the test.
RESTART
RESTART
TEST
TEST
WARNINGS
If a double circuit is not connected properly while the device is performing the last step
of the test (step 6), the device may incorrectly state that a single circuit is installed. If this
happens, reconnect the circuit correctly, press
If the O
sensor is defective (the message "O2 sensor gain calculation: FAILURE" is
2
displayed during the manual test or if the Replace O2 Cell alarm is triggered), do not use
the device until the O
4 Testing the alarms
The operation of all of the visual and audible alarms must be checked regularly, in particular prior
to patient connection and during continuous ventilation.
See "Alarm testing procedures" on page 74.
NOT013281-8
4. Cover the insufflation outlet on the ventilator, then
Cancel
press "Continue". The ventilator measures the rate
of leak, if any.
5. Connect the patient circuit accessories, excluding
the mask (the other end of the patient circuit should
Cancel
be left free), then press "Continue". The ventilator
measures the inspiratory resistance at 30 L/min and
60 L/min.
6. Cover the free end of the patient circuit and press
Cancel
"Continue". The ventilator determines the type of
patient circuit and its resistance and compliance,
and calibrates the expiratory spirometry and the
oxygen sensor.
to return to the main screen. If the test failed, press
Back
Back
sensor has been replaced.
2
, then redo the test.
Back
Back