REMOVING THE FRONT CAP (IF FITTED) ATTACHING YOUR HUMIDAIRE 3i ™ HumidAire 3i ILLUSTRATIONS Component of 33817/4 Front cap FILLING THE WATER CHAMBER Cover Air outlet Cover seal Control dial Connector seal Docking station Heater plate Water chamber Latch REMOVING AND STORING THE CONNECTOR PLUG...
Page 3
STARTING YOUR TREATMENT DISASSEMBLING YOUR HUMIDAIRE 3i REASSEMBLING YOUR HUMIDAIRE 3i Up/Down keys Right key Left key Control panel Start / Stop key 10 min Control dial Indicator light DETACHING YOUR HUMIDAIRE 3i 10 min Large silver button UNDERSIDE VIEW...
Page 4
Manufactured by: ResMed Ltd 1 Elizabeth MacArthur Drive Bella Vista NSW 2153 Australia Tel: +61 (2) 8884 1000 or 1 800 658 189 (toll free) Fax: +61 (2) 8883 3114 Email: reception@resmed.com.au Distributed by: ResMed Corp 14040 Danielson Street Poway CA 92064-6857 USA Tel: +1 (858) 746-2400 or 1-800-424-0737 (toll free) Fax: +1 (858) 746-2900 Email: reception@resmed.com...
UMID HAT IS THE UMID The H 3i is designed to form a single unit with a ResMed CPAP flow UMID generator. The combined unit provides humidification during CPAP therapy, and can make therapy more comfortable. You may have experienced nasal congestion or dryness of the nose, mouth, or throat during CPAP therapy.
EDICAL NFORMATION NTENDED The H 3i is indicated for humidification of the air delivered from the ResMed UMID S8 Series CPAP devices. The H 3i is for use only as recommended by a UMID physician. ONTRAINDICATIONS The H 3i is contraindicated for use with patients whose upper (supraglottic) UMID airway has been bypassed.
P YOUR UMID WARNING Make sure your flow generator is switched off at the power outlet before connecting the HumidAire 3i. EMOVING THE FRONT CAP IF FITTED Please refer to the illustrations in section B of the illustration sheet. Your flow generator may be fitted with a front cap (B-1). Before you can attach the...
Insert the power cord into the socket at the rear of your flow generator. Plug the other end of the power cord into a power outlet (D-3). Note: If your flow generator is powered via the DC input, the HumidAire 3i is not powered and humidification output will be reduced.
CAUTION Do not use any additives (eg scented oils and perfumes). They may reduce the humidification output of the HumidAire 3i and/or cause deterioration of the water chamber materials. Carefully place the filled water chamber into the humidifier docking station (E-5) and latch the cover (E-6 &...
Note: During operation, the indicator light becomes brighter to indicate that the heating element is active. Your HumidAire 3i has a rapid heating element which allows very short heating times. Do not be concerned if the light only brightens very briefly.
Unplug the power cord from the power outlet (G-1). WARNING Take care when handling your HumidAire 3i. The water and the water chamber base may both be hot. Allow ten minutes for the heater plate and any excess water to cool (G-2).
Unplug the power cord from the power outlet (H-1). WARNING Take care when handling your HumidAire 3i. The water and the water chamber base may both be hot. Allow ten minutes for the heater plate and any excess water to cool (H-2).
LEANING YOUR UMID CAUTION Do not use any of the following to clean any part of the HumidAire 3i. These • solutions may cause damage and reduce the life of the product. bleach, chlorine-, alcohol- or ammonia-based solutions • moisturising, antibacterial or glycerine based soaps •...
ResMed. Applicable ResMed warranty details are provided with the device at the time of original supply. Of course, as with all electrical devices, if any irregularity becomes apparent, you should exercise caution and have the device inspected by an authorised ResMed Service Centre.
ROUBLESHOOTING If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the unit. WARNING Service and repair should only be performed by an authorised service agent.
Page 17
No/low air flow The air tubing is not Connect the air tubing as described in connected properly. “Setting Up your HumidAire 3i For Use” on page 3. The air tubing is blocked, Unblock or free the air tubing. Check the pinched or punctured.
ECHNICAL PECIFICATIONS ERFORMANCE Maximum heater plate temperature: 85°C Temperature cut-out: 93°C Maximum output humidity: 95% relative humidity (control dial set to maximum setting, flow rate 50 L/min, ambient temperature of 20°C, ambient relative humidity of 40%). Maximum gas temperature: 41°C IMENSIONS Docking station and water chamber: 118 mm x 170 mm x 212 mm EIGHT...
Page 19
YMBOLS HICH PPEAR ON THE EVICE Attention, consult accompanying documents Type CF equipment Class II equipment Hot surface Do not touch Maximum water level...
Cover seal Power supply 13 cleaning 9 reassembling 9 removing 8 Removing connector plug 4 cover seal 8 Detaching your HumidAire 3i 6 front cap 3 Dimensions 13 water chamber 5 Docking station cleaning 9 Dryness Setting up your HumidAire 3...
Page 21
Warm-Up feature 5 Warnings 2 Water chamber cleaning 9 leaking 11 maximum capacity 13 maximum water level mark 5 Weight 13 Your responsibility 1...