ResMed Elisee 350 Clinical Manual page 64

Table of Contents

Advertisement

Respiratory mechanics parameters
On the curves screen, two screens display the ventilatory mechanics parameters, which can be
accessed using the
Plateau time
Pressing the End-inspiratory pause button inserts a plateau time at the end of the inspiratory
cycle, once the minimum tidal volume has been reached. Hold your finger on the button for the
desired length of the plateau time (max. 5 s). The ventilator will keep the inspiratory and expiratory
valves closed, and no air flow will be delivered.
23
cm
H2O
90
80
70
60
50
40
HP
HP
30
20
10
<PE
<PE
0
26
cm
H2O
90
80
70
60
50
40
HP
HP
30
20
10
<PE
<PE
0
End-expiratory pause
Pressing the End-expiratory pause button inserts a pause at the end of the expiratory cycle.
During the pause, the ventilator keeps the inspiratory and expiratory valves closed.
23
cm
H2O
90
80
70
60
50
40
HP
HP
30
20
10
<PE
<PE
0
27
cm
H2O
90
80
14
70
60
50
28
40
HP
HP
30
20
10
<PE
<PE
0
58/102
(screen 2/4) button and the
2/4
ACV
ACV
M o
O
2
d e
100%
2
s
min
Freeze -> Touch curves
info
P
16
P a
r a
0
m
v
28
0
Pause
Pause
Pause
Pause
Nebul.
Nebul.
E
I
2/4
ACV
ACV
O
2
100%
Back
Back
2
min
Frozen curves !
info
P
16
O
0
v
28
K
0
Ppeak = 26,4 cmH2O
Pplat = 22,4 cmH20
R = 9,3 cmH2O/l/s
ACV
ACV
M o
O
2
d e
100%
2
min
Freeze -> Touch curves
s
info
P
16
P a
r a
0
m
v
28
0
Pause
Pause
Pause
Pause
Nebul.
Nebul.
E
I
2/4
ACV
ACV
O
2
100%
Back
Back
2
Frozen curves !
min
info
P
O
0
v
K
0
PEEPtot= 0,5 cm H2O
PEEPi = 0,1 cmH2O
1. Press screen 2/4
2/4
Pause
Pause
2. Then press
during the inspiratory phase to insert an end-
I
inspiratory pause. The maximum pause time is 5 seconds,
after which expiration is automatically triggered.
The screen displays the plateau pressure (Pplat), maximum
pressure (Ppeak) and resistance (R) values.
3. Press "OK" to go to the frozen curves screen (see "Freezing
the curves" on page 57) or
Note: Wait for the real-time curves to be displayed at the left-
hand side of the screen before pressing the Pause button.
Otherwise, the frozen curves will not display correctly.
1. Press screen 2/4
2/4
Pause
Pause
2. Then press
during the expiratory phase to insert an end-
E
expiratory pause. The maximum pause time is 12 seconds,
after which inspiration is automatically triggered.
The screen displays the total positive expiratory pressure
(PEEPtot) and the intrinsic positive expiratory pressure (PEEPi).
3. Press "OK" to go to the frozen curves screen (see "Freezing
the curves" on page 57) or
Note: Wait for the real-time curves to be displayed at the left-
hand side of the screen before pressing the Pause button.
Otherwise, the frozen curves will not display correctly.
(screen 3/4) button.
3/4
.
to return to the curves screen.
Back
Back
.
to return to the curves screen.
Back
Back
NOT013281-8

Hide quick links:

Advertisement

Table of Contents
loading