ResMed Elisee 350 Clinical Manual page 97

Table of Contents

Advertisement

External power supply ( )
Female power socket FRB DB315
12 to 28 V DC / 15 A max.
RS232 serial link socket (
ResMed-specific equipment.
1: TXD
3
4
2: RXD
3: GND
2
5
4: +5V
5: Digital input/output (0/5 V)
6: NC
6
1
Note: The manufacturer accepts no liability for damage to the device caused by the use of a
cord which does not comply with the recommendations.
Remote alarm (
WARNING
The manufacturer accepts no liability for damage to the device caused by connection to a
remote alarm which does not comply with the specifications provided.
Electrical connection with remote alarm box via specific cord.
Binder 3-pin female Series 712 connector
Note: The ventilator casing is not electrically polarised. It has its own power supply.
Volume level
Volume level: < 40 dBA @ 1 m
When the device is operating on mains power and both buzzers are activated, the volume level is:
86 dBA at a distance of 1 m
78 dBA at a distance of 3 m.
NOT013281-8
2
+
3
-
1
)
)
1 (NC)
3 (NO)
2 (COMMON)
Remote alarm relay:
Max. intensity: 0.25 A
Max. Voltage: 100 V DC connected to ground
or 0 V connected to the device
Ω
Technical specifications
91/102