Active humidification system
A humidification system makes it possible to increase the amount of moisture in the air delivered
to the patient. Set up the heater base and connect it to a power supply in accordance with the
manufacturer's instructions.
Note: If a heated circuit is being used with a humidification system, connect it to a power
supply, then wait 15 minutes before performing a manual test so that the temperature and
humidity can stabilise. This will ensure accurate measurement of the V
circuit).
Water trap (optional)
WARNING
Set up the humidifier in accordance with the manufacturer's instructions. Always ensure
that the ventilator and expiratory valve are placed above the humidifier and water traps.
Note: The water traps must be emptied periodically.
For further details, refer to the user guide for this accessory.
Standard pneumatic nebuliser
The driving gas is pure oxygen with a maximum pressure of 200 kPa and a flow rate of 20 L/min.
Follow the manufacturer's instructions for assembling and connecting the nebuliser.
1
CAUTION
Do not insert a filter between the patient and the nebuliser.
Remote alarm
For further details, refer to the user guide for this accessory.
NOT013281-8
Antibacterial filter
Heated humidifier
2
3
1. Connect the high-pressure oxygen hose.
2. Connect the nebuliser tubing.
3. Connect the nebuliser between the end of the
patient circuit and the patient interface.
(when using a double
TE
Setting up
41/102