ResMed Elisee 350 Clinical Manual page 45

Table of Contents

Advertisement

High-pressure oxygen
High-pressure oxygen can be sourced from:
A wall supply
An oxygen bottle fitted with a pressure regulator. We recommend that you set the regulator at
400 kPa ± 50 kPa.
WARNING
You must only use tubing that complies with the ISO 5359 standard.
With the ventilator in standby mode (the ventilator must not be in ventilation mode):
2
Notes:
We recommend disconnecting all oxygen supplies when the device is on standby or
switched off.
The oxygen bottle can be replaced without stopping ventilation.
Low-pressure oxygen
Low-pressure oxygen can be supplied by:
An oxygen bottle fitted with a pressure regulator and a flow regulator
An oxygen concentrator
Wall-mounted O
CAUTION
Only connect oxygen once ventilation has commenced, and disconnect oxygen before
stopping ventilation.
NOT013281-8
5
6
3
1
fitted with a flow regulator.
2
5. Plug the connector into the sensor.
6. Cover the sensor with the protective cap.
7. Switch the ventilator on, and perform a manual test
to calibrate the sensor (see "3 Manual test" on
page 48). If the message "REPLACE O
not displayed on the screen, the sensor is
operational.
1. Connect the high-pressure oxygen supply to the
O
port on the back of the ventilator.
2
2. If using an oxygen bottle, set the pressure
regulator to 400 kPa ± 50 kPa, then open the
bottle. If the bottle is opened before ventilation
commences, the internal oxygen sensor will be
calibrated to 100% FiO
3. Set the FiO
value.
2
4. Begin ventilation. FiO
40 seconds.
CELL" is
2
.
2
monitoring begins after
2
Setting up
39/102

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents