Adding A Fid, Tcd/Tcd In-Series, Ecd, Or Fpd Detector Module - Thermo Scientific TRACE 1310 Hardware Manual

Gas chromatographs
Hide thumbs Also See for TRACE 1310:
Table of Contents

Advertisement

Adding a FID, TCD/TCD In-Series, ECD, or FPD Detector Module

Thermo Scientific
This section provides instructions for adding a front/back FID, TCD/TCD In-Series, ECD,
or FPD detector module. According to the configuration of your TRACE 1300/TRACE
1310, a dummy module is present into the free site where the detector module is not installed.
See
Figure
446.
Figure 446. Add a Front/Back Detector Module
Dummy Front Detector Module
ATTENTION Where a dummy module is installed, the gas connection is blocked by a
plug.
To add a new FID, TCD/TCD In-Series, ECD, or FPD detector module
1. Put the GC in standby condition.
2. Cool the oven, injector, and detector to room temperature.
Note By pressing the Maintenance button, the GC cool down is automatically
carried out.
3. Close the gas supplies.
4. Power off the GC.
a. Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
5. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
Adding a FID, TCD/TCD In-Series, ECD, or FPD Detector Module
Dummy Back Detector Module
TRACE 1300 and TRACE 1310 Hardware Manual
9
Adding Modules
431