Connecting The Oven Cryogenic System; Oven Cryo System With Carbon Dioxide Connection - Thermo Scientific TRACE 1310 Hardware Manual

Gas chromatographs
Hide thumbs Also See for TRACE 1310:
Table of Contents

Advertisement

Connecting the Oven Cryogenic System

Oven Cryo System with Carbon Dioxide Connection

Thermo Scientific
If the oven of your TRACE 1300/TRACE 1310 is equipped with a cryogenic cooling option,
you need a supply of coolant, such as carbon dioxide (CO
This Oven cryo system consists of a tube for carbon dioxide inserted into the oven and fitted
with a 1/8-in. solenoid valve for carbon dioxide mounted into the back of the GC.
Carbon dioxide must be supplied in a high-pressure cylinder with a dip tube. It is your
responsibility to ensure the delivery connection from the carbon dioxide cryogenic supply is
adaptable to 1/8-in. tubing.
configuration.
Figure 13. Carbon Dioxide Container
CORRECT
Dip tube
CAUTION High pressures and extremely low temperatures make pressurized carbon
dioxide a hazardous material. High concentrations of carbon dioxide are dangerous.
To avoid injury, always follow the safety precautions and delivery system design
recommended by your gas supplier.
To connect the cryogenic system to the carbon dioxide supply
1. Connect the proper end of the cryo supply tube to the 1/8-in. NPT connection of the
solenoid valve using the proper nut and ferrule. Use a 7/16-in. wrench to tighten the
fittings. See
Figure
14.
Figure 13
shows the proper carbon dioxide container
WRONG
TRACE 1300 and TRACE 1310 Hardware Manual
1
Connecting the Oven Cryogenic System
) or liquid nitrogen (N
2
2
Installation
).
13

Advertisement

Table of Contents
loading

This manual is also suitable for:

Trace 1300

Table of Contents