Upgrading A Stand Alone Trace 1300/Trace 1310 To Ms Version; Preliminary Operations; Getting Started - Thermo Scientific TRACE 1310 Hardware Manual

Gas chromatographs
Hide thumbs Also See for TRACE 1310:
Table of Contents

Advertisement

11
Upgrade Equipment

Upgrading a Stand Alone TRACE 1300/TRACE 1310 to MS Version

Upgrading a Stand Alone TRACE 1300/TRACE 1310 to MS Version

Preliminary Operations

Getting Started

556
TRACE 1300 and TRACE 1310 Hardware Manual
This section provides the instructions for updating your TRACE 1300/TRACE 1310 stand
alone version to a MS version. According to Thermo Scientific you need a dedicated upgrade
GC
kit to couple the
with an ISQ Series, TSQ 8000 Series, DSQ II, ITQ, or TSQ Quantum
mass spectrometer.
WARNING This operation must be carried out by authorized and trained Thermo Fisher
Scientific Service Field Engineers.
Before starting, the following preliminary operation must be carried out.
1. Put the GC in standby condition.
2. Cool the oven, injectors and detectors to room temperature.
3. Close the gas supplies.
4. Power off the GC.
a. Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
5. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
Depending on your mass spectrometer, see the following sections:
Coupling with the ISQ Series / TSQ 8000 Series Mass Spectrometer
Coupling with the DSQ II Mass Spectrometer
Coupling with the ITQ Mass Spectrometer
Coupling with the TSQ Quantum Mass Spectrometer
IMPORTANT The temperature control for the DSQ II, ITQ and TSQ Quantum transfer
line is carried out through the Aux Temperature/Cryo Module. See the section
an Aux Temperature/Cryo Module"
on
page 443
for details.
"Adding
Thermo Scientific

Advertisement

Table of Contents
loading

This manual is also suitable for:

Trace 1300

Table of Contents