4. Trouble Shooting

4.1 Error Messages: Problem / Solution

The error messages are listed to help localize possible errors faster.
The possible error referred to in this chapter may not always be the case, as they are only theoretically
occurring errors and solutions.
Always keep us informed about any kind of error occurring, which is not listed in this chapter. With this
information provided, we are able to improve and complete this operation manual.
Many thanks in advance for your support.
HERMLE Labortechnik GmbH

4.2 Survey of Possible Error Messages and Solutions

4.2.1 Lid Release during Power Failure (Emergency Lid Release)

In case of power failure or malfunction, the lid of the centrifuge can be opened manually, in order to
protect samples.
Please proceed as follows:
• Switch the centrifuge off and unplug the power cord, wait until the rotor is at a standstill,
(this may take several minutes).
At the right side of the centrifuge, there is a plastic stopper (Figure 35). Remove this stopper, which
is connected to the lid lock, horizontally from the housing until the centrifuge lid opens.
• Now, open the lid of the centrifuge.
Figure 35
© Hermle Labortechnik GmbH
Z307_V1.22_eng
TROUBLE SHOOTING
27