Trouble Shooting; Error Message: Problem / Solution; Survey Of Possible Error Messages And Solutions; Lid Release During Power Failure (Emergency Lid Release) - Hermle Z 366 K User Manual

Table of Contents

Advertisement

MAINTENANCE

4. TROUBLE SHOOTING

4.1 Error Message: Problem / Solution

The error messages are listed to help localize possible errors faster.
The possible error referred to in this chapter may not always be the case, as they are only theoretically
occurring errors and solutions.
Always keep us informed about any kind of error occurring, which is not listed in this chapter. With this
information provided, we are able to improve and complete this operation manual.
Many thanks in advance for your support.
HERMLE Labortechnik GmbH

4.2 Survey of Possible Error Messages and Solutions

4.2.1 Lid Release during Power Failure (Emergency Lid Release)

In case of power failure or malfunction, the lid of the centrifuge can be opened manually, in order to
protect samples.
Please proceed as follows (see Figure 33):
• Switch off the centrifuge and unplug the power cord, wait until the rotor has come to a
standstill (this may take several minutes)
• On the left hand side of the centrifuge housing, there is a plastic stopper. Remove this
stopper and behind is a hexagon nut.
• Take the delivered box spanner, put it into the hole, and lock the box spanner, with the
hexagon nut (see Figure 31).
• Turn the box spanner to the right (clockwise), up to the limit.
ATTENTION: Only turn to the limit, don´t tighten the nut.
• Open the lid of the centrifuge.
• Switch the centrifuge on again, to proceed with regular function.
Figure 33
34
Z366K_V1.22_eng
© Hermle Labortechnik GmbH