9.1.1 Data available

All individual data are separated by at least one tabulation character (ASCII code=0x09).
For the cyclic measurements, the data format is detailed. For the files, only one example for
each file is given to explain the data format.
Cyclic measurements
1. If the option "Mode = standard" is chosen, the following message is sent:
CHn\t Gas\t Gas Unit\t Temperature\t Temperature Unit\t Barometric Pressure\t Barometric Pressure Unit\t\ Event\t\r\n
with:
\t ................................................the ASCII tab character: code=0x09
\r ................................................the ASCII Carriage Return character: code=0x0D
\n ..............................................the ASCII Line Feed character: code=0x0A
CHn ..........................................the 2 ASCII characters "CH" + the channel no.
Gas ............................................the gas concentration.
Gas Unit ....................................the gas unit.
Temperature ..............................the temperature.
Temperature Unit.......................the temperature unit.
Barometric Pressure..................the barometric pressure.
Barometric Pressure Unit ..........the barometric pressure unit.
Event .........................................the event bit mask in hexadecimal format.
The values are not described here. Refer to
Example of one measurement:
CH1
2. If the option "Mode = expert" is chosen, the following message is sent:
CHn\t Gas\t Gas Unit\t Temperature\t Temperature Unit\t BarometricPressure\t Barometric Pressure Unit\t
with:
Phase shift.................................The fluorescence phase shift in [°].
Partial pressure .........................The partial pressure in [bar].
Time ..........................................The time of the measurement. Format "hh:mm:ss."
Index..........................................This is the index of the last measurement.
This number starts at 0 at power up of the program. The following is an example of one
measurement:
CH1
bar
5923
697.176
mbar 20.1
Event\t Phase Shift\t °\t Partial Pressure\t bar\t Reference Phase\t °\t
Fluorescent Phase\t °\t Reference Amp\t V\t Fluorescent Amp\t V\t
Tinst\t Tinst Unit\t Offset\t Time\t Index\r\n
697.173
ppb
20.1
-21.409
°
-64.991
List of events and alarms on page
°C
0.982
bar C00
°C
0.982
bar
03000000
°
2.349
V
2.499
Communication Menu
100.
26.045
°
V
25.531 °C
45.000
0.69700
22:59:42
75