Code Value Short description
error
116
[none] Type: ADC=23, Timeout=231,
[none] Channel
[none] Sampling time
standard
117
message
[none] Mainboard serial
[none] Bootloader version
[none] Hardware version
[none] Module ID
[none] BLE module version (if equipped)
[%]
[none] HWC LED adjust timestamp
[none] BLE module MAC address (if
error
118

6.2 Get the data logger

Error: ADC
Invalid=232
Instrument information
HWC LED adjust
equipped)
Parameter load error
1. Install the service cable VAA10036.99.G00 in the DR300. Refer to
cable
on page 19.
2. Start the VAP235 software.
3. Connect the service cable VAA10036.99.G00 to an USB port.
4. Push
on the DR300.
5. Push Connect. The display of the DR300 shows serial.
6. Push Get data log.
7. Define the path to store the data logger. Refer to
8. Find the csv file in the path which is defined. E.g.
DL_LPG445.99.00000_SN18060A000047_180926_1544.csv. Refer to
Detailed description
[DR300]ADC_CHANNEL_LED_VOLTAGE=1
ADC_CHANNEL_LED_MON=2
ADC_CHANNEL_BATTERY_VOLTAGE =3
ADC_CHANNEL_HARDWARE_ID=4
ADC_CHANNEL_MODULE_ID_CODE =15
ADC_CHANNEL_STM32_TEMPERA TURE=17
Figure
13.
Troubleshooting
Insert the service
Figure
14.
61