Configuration report example
GENERAL CONFIGURATION
Measurement mode . . . . . . . . . Continuous mode
Pressure unit . . . . . . . . . . . . . . . bar
Temperature unit . . . . . . . . . . . . °C
Storage mode . . . . . . . . . . . . . . Rolling buffer
Storage RAM time . . . . . . . . . . . 10 [s]
Storage FLASH time . . . . . . . . . 3600 [s]
Autosave in FLASH. . . . . . . . . . Disabled
Channel 1
Sensor. . . . . . . . . . . . . . . . . . . . POWER OT
Medium . . . . . . . . . . . . . . . . . . . Liquid
Gas unit. . . . . . . . . . . . . . . . . . . ppm->ppb
Liquid . . . . . . . . . . . . . . . . . . . . Water
Resolution displayed . . . . . . . . . 1
Thermal alarm . . . . . . . . . . . . . . Enabled 90.0 [°C]
Measurement interval . . . . . . . . 5 [s]
Out of range protection . . . . . . . Enabled
User offset. . . . . . . . . . . . . . . . . Enabled
User offset value . . . . . . . . . . . . 45
Alarms
Low Low . . . . . . . . . . . . . . . . . . Enabled
Low . . . . . . . . . . . . . . . . . . . . . . Enabled
High. . . . . . . . . . . . . . . . . . . . . . Enabled
High High . . . . . . . . . . . . . . . . . Enabled
Hysteresis . . . . . . . . . . . . . . . . . 5 [%]
Delay . . . . . . . . . . . . . . . . . . . . . 15 [s]
Filter
State . . . . . . . . . . . . . . . . . . . . . Disabled
Type . . . . . . . . . . . . . . . . . . . . . Median
Depth . . . . . . . . . . . . . . . . . . . . 5
Central depth . . . . . . . . . . . . . . 0

9.1.2 Example of use

In this example we use:
One PC with a RS232 port.
One "RS-485<->RS232 converter"
Procedure:
1. Connect both RS-485 wires of the instrument to the "RS-485<->RS232 converter".
2. Connect the "RS-485<->RS232 converter" to the PC RS232 port using a standard cable
(RS232 DB9 straight cable).
On the PC:
1. Run "Hyperterminal" on the PC.
2. Configure the PC COM port used (e.g. COM2). Menu "File/Properties/Configure".
3. Configure the parameters "Baud rate", "Parity", "Nb of stop bits", "Byte size" (Menu
"File/Properties/Configure"). Use the same parameters for the instrument and the PC.
4. Configure the "Font = Courier 10" (Menu "View/Font").
5. Connect "Hyperterminal" (Menu "Call/Call").
6. Save the data received in the file of your choice (Menu "Transfer/Capture Text/Start").
On the instrument:
1. Use the menu "Communication/RS-485 Simple/Send files" and the button "All files".
When the transfer is finished, close the file with "Hyperterminal" (Menu "Transfer/Capture
Text/Stop"). Now, all the reports are saved in a text file on your PC.
Communication Menu
100.000000 [ppb]
200.000000 [ppb]
9000.000000 [ppb]
10000.000000 [ppb]
77